Maximum number of verification intervals: nLC = 3000
Full Scale Output: 2mV/V +/- 10%
Temperature effect on zero:
from ±0,0093% F.S./10K to ±0,0140% F.S./10K
Temperature effect on full scale output:
±0,0117% F.S./10K (-10°C/+20°C)
±0,0175% F.S./10K (+20°C/+40°C)
Compensated Temperature Range: -10°C/+40°C
Operating Temperature Range: -10°C/+50°C
Creep at nominal load in 30 minutes: ± 0,01% F.S.
Maximum tolerated excitation voltage: 15 VDC
Input Resistance: 380 ± 15 Ohm
Output Resistance: 359 ± 5 Ohm
Insulation Resistance at 100V: >1000 MOhm
Safe Overload: 150% F.S.
Ultimate Overload: 300% F.S.
Nominal displacement: < 0.5 mm
6-wire shielded cable, 3m long, 5mm diameter
The CCATEX certificate makes the load cell suitable for use in hazardous areas with protection methods according to:
ATEX II 1G Ex ia IIC T6 (Ta -20÷+40°C) TX (Ta -20÷+65°C) Ga
ATEX II 1D Ex ta IIIC TX°C (Ta -20÷+40°C) TX°C (Ta -20÷+65°C) Da IP65.
The certificate includes the manual and the ATEX CE conformity declaration (available in EN, DE, FR, and IT).
IMPORTANT: the certificate refers to the single load cell. For systems with more load cells, it is necessary to order one certificate for each load cell.